Vicks V980CA Uživatelský manuál Strana 9

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 20
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 8
9
Product Specifications
Displayed temperature range: 35.5–42 °C (95.9–107.6 °F)
Measurement time: 1 second
Operating ambient temperature range: 10–40 °C (50–104 °F)
Display resolution: 0.1 °C (also available in °F)
Accuracy:
Non-predictive mode (liquid bath):
37 - 39 °C 0.1 °C (98.6 - 102.2 °F 0.2 °F)
below 37 & above 39 °C 0.2 °C (below 98.6 & above 102.2 °F 0.3 °F)
Predictive mode (clinical) : 0.2 °C (0.4 °F)
Humidity storage: < 15-95% non-condensing
Battery life: 2 years or 500 measurements
Symbol Explanations:
Subject to change without notice.
This appliance conforms to ASTM Standards for Clinical Thermometers:
E1112 and E1965-98 for clinical accuracy, wherever applicable.
Portable and mobile RF communications equipment can affect MEDICAL
ELECTRICAL EQUIPMENT.
This product conforms to IEC60601-1 (General Requirements for Safety) and
IEC60601-1-2 (Electromagnetic Compatibility)
LOT
TYPE BF
EQUIPMENT
LOT
NUMBER
DATE OF